Insulin Pump Recall: Tandem Diabetes Removes Tandem Mobi Insulin Pumps
The FDA is aware that Tandem Diabetes has issued a letter to affected customers recommending certain insulin pumps identified for replacement be removed from where they are used or sold. Tandem Diabetes also issued a separate letter providing updated use instructions for all other Tandem Mobi Insulin Pumps with Interoperable Technology.
Tandem Diabetes stated that certain Tandem Mobi insulin pumps may generate false-positive occlusion alarms when no actual insulin flow obstruction is present. As a result, the pump may incorrectly detect an occlusion condition and suspend insulin delivery. When an occlusion alarm occurs, insulin delivery remains suspended until the user acknowledges the alarm and resumes therapy. Repeated false-positive occlusion alarms, if not addressed promptly, may result in interruptions of insulin delivery that can increase the risk of elevated blood glucose levels. In severe cases, hyperglycemia may require medical intervention or hospitalization.
As of October 28, 2026, Tandem Diabetes has reported 1,248 serious injuries and no deaths associated with this issue.
Questions?
Customers in the U.S. with questions about this recall should contact Tandem Diabetes at techsupport@tandemdiabetes.com or 877-801-6901.
This recall involves removing certain devices from where they are used or sold. This recall also involves updated use instructions for devices not subject to removal. The FDA has identified this recall as the most serious type. This device may cause serious injury or death if you continue to use it.
|