Continuous Glucose Monitoring Software Correction: Dexcom Issues Correction for Dexcom G6 and G6 Pro Software
Dexcom Inc. is correcting the Dexcom G6 Continuous Glucose Monitoring System’s G6 and G6 Pro Android US CGM App version 1.15.0 due to an identified bug that can cause the app to terminate unexpectedly. As a result, the user may not receive estimated glucose values, alarms, or alerts. If a user is unaware that the app has terminated, there is potential for missed detection of a high blood sugar (hyperglycemic) or low blood sugar (hypoglycemic) event.
The use of affected product may cause serious adverse health consequences, including hyperglycemia, hypoglycemia and death.
This recall involves correcting devices and does not involve removing the devices from where they are used or sold. The FDA has identified this recall as a Class I recall, the most serious type. This device may cause serious injury or death if you continue to use it.
Questions?
Customers in the U.S. with questions about this recall should contact Dexcom Technical Support at 1-844-478-1600.
|
|
|
|