Type: Ventilator, continuous, facility use
Manufacturer: Maquet Critical Care AB Brand: Servo-U Model: Servo-U
Event: A Biomedical Engineering team member noticed the side arm that holds a humidifier on the ventilator was cracked near the center hinge. The team checked all Getinge Servo-U ventilators and found that two ventilators had the same cracking at the metal arm or were beginning to show signs of cracking in the identical location. Another 43 ventilators showed scratches in the same location. The manufacturer was notified and has expedited replacement arms. They have not responded with a weight rating for the ventilator arm.
Photo: Scratch present on the ventilator arm, the same location a crack was found.
|
Type: Catheter, intravascular, therapeutic, long term greater than 30 days
Manufacturer: Bard Access Systems Brand: BD Bard PowerPICC Solo 4FR Model: 3194355 Ref: 3194355
Event: The patient originally had a peripherally inserted central catheter (PICC) line placed for antibiotics to be given and was discharged from the hospital at that time. The patient was readmitted one month later, and it was noted the PICC site was swollen. Doctor orders were placed to remove the PICC line and send the tip for culture. During the removal, the nurse documented that there was no resistance or difficulty felt when removing the line. However, only a portion of the PICC line came out. The doctor was immediately notified, and a chest and arm X-ray was obtained along with an ultrasound of the arm to verify that the retained piece was still in the arm. The patient was taken to interventional radiology and the retained piece was removed. The device was shipped to the manufacturer for review.
Type: Pump, infusion
Manufacturer: B. Braun Medical Inc. Brand: Infusomat Space S UDI-DI: 04046964660887 Cat: 9713051U, 8713051U
Event: Patient was on multiple critical medications given through a catheter (IV pump). The IV pump alarmed low battery even though it was plugged into an emergency outlet and the charging light was lit. A backup pump was obtained and started. Approximately one hour later, the new IV pump alarmed low battery even though it was plugged in. Then, the pump abruptly shut off within one minute of alarming. The patient’s blood pressure dropped quickly and required rapid titrations once a new IV pump was set up.
Type: Set, administration, intravascular
Manufacturer: Baxter Healthcare Corp Brand: Duo-Vent UDI-DI: 00085412676630 Model: 2R8404 Lot: DR24C22079 Ref: 2R8404
Event: The dark blue slide clamp key is on backwards and you must twist the tubing to use it. This makes it challenging to deliver chemotherapy through the catheter.
Photo: The red circle and arrow indicate where the slide clamp key inserts into the pump.
|
Type: Set, administration, intravascular
Manufacturer: Baxter Healthcare Corp Brand: Duo-Vent UDI-DI: 0085412676630 Model: 2R8404 Ref: 2R8404
Event: A nurse was placing the catheter to deliver the chemotherapy agent Pemetrexed (Alimta) into the patient. When the blue key was removed from the pump, it sliced the tubing and chemotherapeutic agent spilled from the bag onto floor and the base of the pump.
Type: General surgery tray
Manufacturer: DeRoyal Industries, Inc. Brand: DeRoyal Laparotomy Pack UDI-DI: 00749756364430 Model: 89-10558 Cat: 89-10558 Lot: 61045501
Event: An unknown substance on the syringe plunger was found when staff opened a sterile pack. The substance was identified prior to patient contact.
|
|