Syringe Recall: Cardinal Health Removes Certain Monoject U-100 mL Insulin Syringe Luer Lock with Tip Cap Soft Packs Due to Incompatibility with Needleless Connectors
Cardinal Health is recalling certain lots of Monoject U-100 1 mL Insulin Syringe Luer-Lock with Tip Cap Soft Pack due to an incompatibility with needleless IV connectors. If the affected device is used to administer IV push insulin through a needleless connector, it may increase the risk that the patient may not receive the full dose of insulin because not all the medication will leave the syringe.
The FDA has identified this as a Class I recall, the most serious type of recall. This recall involves removing certain devices from where they are used or sold. This device may cause serious injury or death if you continue to use it.
Questions?
Customers in the U.S. with questions about this recall should contact the appropriate Cardinal Health customer service group: Hospital – 1-800-964-5227, distributor – 1-800-635-6021, all others – 1-888-444-5440.
|
|
|
|