Cincinnati, OH – May 10, 2024 – MEGADYNE™ is initiating a voluntary recall of all MEGADYNE™ MEGA SOFT™ Pediatric Patient Return Electrodes. The product is being removed because Megadyne has received reports of patient burn injuries in procedures where the Mega Soft Patient Return Electrodes were used. A root cause investigation on the reports included testing which showed a combination of factors when present together may result in potential for thermal injuries.
The combination of these conditions may be more likely when the pad is used with infants and small children. Because the pediatric pad is designed for patients between 0.8 to 50 pounds, which would be predominantly patients under the age of 12, the decision was made to discontinue and recall the 0840 pediatric pad product.
|
|
|
|