Tandem Diabetes Care, Inc. is recalling version 2.7 (released February 12, 2024 for devices using the Apple iOS platform) of the t:connect mobile app used in conjunction with t:slim X2 insulin pump with Control-IQ technology. This recall is a correction and not a product removal. The reason for the recall is due to an issue with the software that may cause the mobile app to crash and be automatically relaunched by the iOS operating system. This cycle intermittently repeats, which leads to excessive Bluetooth communication that may result in pump battery drain and may lead to the pump shutting down sooner than typically expected.
The FDA has identified this as a Class I recall, the most serious type of recall. Use of these devices may cause serious injuries or death. Please be aware, this recall is a correction, not a product removal.