Becton Dickinson (BD)/Carefusion 303 Recalls Alaris Infusion Pumps due to compatibility issues with Cardinal Health Monoject syringes. Use of incompatible syringe sizes and models with the BD Alaris Syringe and PCA Modules can impact syringe pump operation resulting in inaccurate fluid delivery, delayed generation of occlusion alarms, and other potential problems.
The FDA has identified this as a Class I recall, the most serious type of recall. Use of these devices may cause serious injuries or death. Please be aware, this recall is a correction, not a product removal.
Questions?