Hamilton Medical Inc. Recalls HAMILTON-C1, C2, C3, T1 Ventilators for Software Issues that May Cause Ventilators to Stop Without Notice
Hamilton Medical Inc. is recalling the HAMILTON-C1, C2, C3, T1 ventilators after receiving reports of software issues that may cause the ventilators to stop without notice. The ventilator will switch to 'Ambient State,’ if the device is used for more than 91 days without a restart.
There have been 80 reported complaints, no injuries and no deaths related to this recall.
The FDA has identified this as a Class I recall, the most serious type of recall. Use of these devices may cause serious injuries or death.