GE HealthCare is recalling malfunctioning TruSignal sensors that may reduce the amount of energy sent to the heart during defibrillation without any notification to the care provider, which could prevent delivery of lifesaving therapy in a critical situation. This issue is most hazardous to hospitalized patients who may need defibrillation for cardiac arrest. Affected sensors may also unintentionally expose patients to electrical currents from other sources or may provide inaccurate measurements of SpO2, which can impact treatment decisions.
The U.S. Food and Drug Administration (FDA) has identified this as a Class I recall, the most serious type of recall. Use of these devices may cause serious injuries, serious health consequences, or death.
Questions?
Customers with questions or concerns should contact GE HealthCare Service at 1-800-437-1171 or their local service representative.
|
|
|
|