If the user encounters the software issue where the graphical Biopsy Depth Gauge is no longer synchronized with other navigation views, it may result in a prolonged procedure, the need for an additional surgical procedure, aborted procedure, tissue injury, including potential for life-threatening injury (such as hemorrhage, unintended tissue damage, or permanent neurological injury), which could lead to death.
There have been four complaints regarding this device issue. There have been no reported injuries or deaths.